Are you looking for a quote on 6640NCM104377? If so, please fill out the provided quote form below. The NSN (national stock number) is 6665015852354, and the part 6640NCM104377 is radiacmeter by Ludlum Measurements Inc. Please enter the given detail as much as possible, this information will help us to provide you the right quote to fulfill your needs. We will get back to you in less than 15 minutes!
MRC | Criteria | Characteristic |
---|---|---|
ABHP | OVERALL LENGTH | 8.500 INCHES NOMINAL |
APPJ | RADIATION RANGE PER HOUR | 0.2 TO 50000.0 MILLIROENTGENS |
AMRM | RADIATION TYPE AND FUNCTION | BETA RAYS MEASURES GAMMA RAYS MEASURES |
ABKW | OVERALL HEIGHT | 9.200 INCHES NOMINAL |
ABMK | OVERALL WIDTH | 3.500 INCHES NOMINAL |
FEAT | SPECIAL FEATURES | RUGGED AIR IONIZATION CHAMBER; PERFORMS BETA-GAMMA DOSE RATE MEASUREMENTS OVER A FIVE-DECADE SPAN RANGING FROM BACKGROUND TO 500 MSV/HR (50 R/HR); CHAMBER VOLUME: 13.4 CU IN; WINDOW: 7 MG/CM2 ALUMINIZED MYLAR; WINDOW AREA: 6.2 IN2; BETA SHIELD: RETRACTABLE 1000 MG/CM2 PHENOLIC SLIDE; SIDE WALL: 1000 MG/CM2 ALUMINUM AND ACRYLIC; SIX-POSITION SWITCH SELECTS OFF, X10K, X1K, X100, X10, AND X1; EACH RANGE HAS FRONT-PANEL MOUNTED CALIBRATION CONTROL; ALSO INCLUDES AUDIO OUTPUT, RESET, AND ZERO ADJUST CAPABILITIES; ALSO HAS EANS TO ZERO THE ELECTROMETER LEAKAGE CURRENT; APPROX. WT: 3.6 LBS |
CXCY | III PART NAME ASSIGNED BY CONTROLLING AGENCY | ION CHAMBER MODEL 9-3 LUDLUM |
Are you looking for quote within 15 minutes for your desired part? Fill an instant form and receive lowest price quote guaranteed.